Year : 2016  |  Volume : 36  |  Issue : 1  |  Page : 6-13

Accuracy of spinal screw fixation using intraoperative O-arm navigation: Consecutive series of 118 screws


Department of Neurological Surgery, National Defense Medical Center, Tri-Service General Hospital, Taipei, Taiwan, Republic of China

Correspondence Address:
Yuan-Hao Chen
No. 325, Section 2, Cheng-Kung Road, Neihu 11490, Taipei, Taiwan
Republic of China
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/1011-4564.177169

Rights and Permissions

Background: Imaging-guided surgery provides intraoperative realtime navigation for spinal surgery to prevent neurovascular injury. However, the initial experience of O-arm three-dimensional (3D) navigation is less clarified. The study aims to evaluate the accuracy of O-arm 3D navigation-assisted spinal implantation. Materials and Methods: Total 118 transpedicle/lateral mass screws in 17 consecutive patients performed through O-arm 3D computed tomography (CT) imaging navigation. Screws accuracy, Visual analogue scale (VAS), and operation time were assessed. O-arm 3D navigation included 96 pedicle screws and 22 lateral mass screws. Results: There accuracy rate of screw implantation was 96.6% (114/118) without breach (Grade 0), whereas 3.4% (4/118) breach between 2 and 4 mm (Grade 2). In the cervical spinal, 12 pedicle screws were placed in 4 patients; 16.7% (2/12) exhibited a Grade 2 breach in one patient, who developed a new neurological deficit and required revision. In the lumbosacral spine, 2.78% (2/72) exhibited a Grade 2 breach in 2 patients. The mean ± standard deviation VAS of the patients in postoperative and preoperative status was 1.47 ± 0.50 and 3.58 ± 1.00, respectively. The operation time was significantly longer in O-arm navigation than in C-arm guidance (426.5 ± 180.4 vs. 317.9 ± 133.6 min, P < 0.05). Conclusion: O-arm 3D navigation achieves a relatively high accuracy of pedicle and lateral mass screws implantation. The accumulation of experience for O-arm 3D CT during initial learning curve is still warranted to promote the accuracy of screws position and shorten operation time.


[FULL TEXT] [PDF]*
Print this article     Email this article
 Next article
 Previous article
 Table of Contents

 Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
 Citation Manager
 Access Statistics
 Reader Comments
 Email Alert *
 Add to My List *
 * Requires registration (Free)
 

 Article Access Statistics
    Viewed3019    
    Printed46    
    Emailed0    
    PDF Downloaded269    
    Comments [Add]    
    Cited by others 2    

Recommend this journal